MATERIAL
|
DESCRIPTION
|
CONCERN
|
REFERENCE
|
Breathing system equipment (built-in)
|
Supply hose with door connection, demand valve mask, face tent, trach collar, non-rebreather masks
|
Comfort, Pharmacologic
|
Midwest Hyperbaric Operations Manual
|
Catheter, (indwelling urinary cath or wound drainage system)
|
Flexible tubing with collapsible reservoir Empty excess drainage, purge air
|
Comfort, Pressure
|
Midwest Hyperbaric Policy & Procedure Manual
Midwest Hyperbaric Operations Manual
|
Cloth restraints
|
No velcro or metal components, use kerlix roll for restraint
|
Energy, Fire, Acrylic Damage
|
Midwest Hyperbaric Policy & Procedure Manual
Midwest Hyperbaric Operations Manual
|
Contact lenses (soft and gas permeable)
|
Comfort
|
|
|
Cotton gauze dressings
|
Pads, packing, wraps, island-type with non-stick surface
|
Comfort, Pharmacologic
|
|
Diaper (adult/infant) / Kotex
|
Should remain against patient skin, moist, no velcro closures
|
Comfort, Infection
|
Midwest Hyperbaric Policy & Procedure Manual
|
Dressings (aqueous) or topical medication
|
Hydrocel
|
Pharmacologic, Fire, Comfort
|
Midwest Hyperbaric Policy & Procedure Manual
|
ECG and pressure line monitoring cables
|
Biomed approved/maintained
Use of pass-through to connect to monitor outside chamber.
|
Energy
|
NFPA 99 14.2.7.6 [1]
|
Eyeglasses
|
Plastic or metal composition, lenses
|
Comfort
|
Midwest Hyperbaric Safety Discussion
|
Feeding tube (Dobhoff)
|
Flushed and clamped
|
Comfort, Pressure, Pharmacologic
|
Midwest Hyperbaric Policy & Procedure Manual
|
Gowns, hospital issue
|
100% cotton, unit supplied, pocket less, no metal snaps or Velcro, uncontaminated.
Must check for concealed items
|
Comfort, Fire
|
NFPA 99 14.3.1.5.4
[1]
|
Heimlich Valve
|
Attached to indwelling chest tube and sterile glove or urinary drainage bag
|
Pharmacologic, Comfort, Pressure
|
Midwest Hyperbaric Operations Manual
|
IV extension tubing for Monoplace Hyperbaric chamber
|
Use of pass-through to connect to hyperbaric IV pump outside chamber
|
Pressure
|
Midwest Hyperbaric Operations Manual
Midwest Hyperbaric Policy & Procedure Manual
|
Mattress, wedges, pillows
|
Only those recommended by chamber manufacturer
|
Comfort, Fire, Pressure
|
Midwest Hyperbaric Policy & Procedure Manual
|
Nail polish, acrylic nails |
Allowable if applied no less than 12 hours prior to treatment |
Fire |
Midwest Hyperbaric Policy & Procedure Manual |
Nasogastric tube
|
Un-clamped with end loosely covered
|
Comfort, Pressure
|
Midwest Hyperbaric Policy & Procedure Manual
|
Nebulizer equipment (in-chamber)
|
Disposable nebulizer, 5-in-1 adapter, liquid medication
|
Pharmacologic, Comfort
|
Midwest Hyperbaric Policy & Procedure Manual
|
Plastic drink cup
|
No-spill style, vented, unit supplied, only approved unit-supplied fluid
|
Comfort, Pressure, Fire, Pharmacologic
|
Midwest Hyperbaric Safety Discussion
|
Plastic tape
|
No paper or silk tape
|
Fire, Energy
|
NFPA 99 14.3.1.5.4.1 [1]
|
Pressure infuser with spacer
|
Ethox Surgi-Press, oxygen-safe lubricant (silicon base)
|
Pressure, Pharmacologic
|
NFPA 99 14.3.6.2.1.4 [1]
|
Sheets, pillow cases, towels, wash cloths, blankets
|
100% cotton, unit supplied
|
Comfort, Fire
|
NFPA 99 14.3.1.5.4.2 [1]
|
TCOM electrode with cord
|
Radiometer – Max 28v and 4w in class B chamber Use of pass-through to connect to monitor outside chamber
|
Energy
|
NFPA 99 14.2.7.6.1, 14.2.7.6.1.1 [1]
|
Urinal, bedpan
|
Disposable, plastic
|
Comfort
|
|
Ventilator breathing circuit (hyperbaric)
|
Sechrist Model 500A
Use of gas port to connect to control module outside chamber
|
Pressure, Pharmacologic
|
Midwest Hyperbaric Operations Manual
Midwest Hyperbaric Policy & Procedure Manual
|