Chamber: Single
More suitable alternative: NO
Produces heat: No Info
Risk of Fire and Explosion - Probability: 1 x Exposure: 4 x Consequence: 3 = Risk Score: 12
Produces Static: No Info
Risk of Fire and Explosion - Probability: 3 x Exposure: 4 x Consequence: 3 = Risk Score: 36
It is a synthetic material, it will be inside the fiberglass cast and the cast will be grounded by lying directly on Sechrest mattress therefore it will be grounded as the matress is conductive.
Low Flash point: NO
Produces Fuel: NO
Produces Adverse: NO
Produces Mechanical: NO
See table below to understand risk level corresponding to each risk score.
Risk Score Risk Level Risk Description
> 100 5 Extremely dangerous: do not operate facility
50 - 100 4 Very high: stop use of equipment, processs or procedure
20 - 50 3 High: required urgent attention; proceed with care
5 - 20 2 Medium: attention needed but operation may continue
< 5 1 Low: acceptable risk but noted
* Burman F. Risk Assessment Guide for Installation and Operation of Clinical Hyperbaric Facilities. San Antonio, Texas: International ATMO; 6th Edition, 2019. ISBN: 978-0-578-22098-7

Created 6/6/2024 6:26:48 PM by FMC HBOT   Mediawebserver 3M Stockinet.pdf
t
-->